DKI Jakarta - Jakarta Selatan

OTDR EXPO FTB-730

Harga :
CALL
Merk :
EXFO
Pembelian Minimal :
1 Unit
Pembelian Maksimal :
10 Unit
Pengiriman :
1 hari sejak pembayaran
Lokasi :
DKI Jakarta - Jakarta Selatan
Dilihat :
239 Kali
Update Terakhir :
03-06-2015 14:56

Detail Produk

Optimizing Passive Optical Network (PON) Testing
Designed to meet testing requirements brought by FTTH networks in general, and PONs/MDUs in particular, the FTB-730 enables testing at 1310, 1490, 1550 and 1625/1650 nm. What's more, EXFO's next-generation OTDR software lets you test through high-port-counter splitters—even 1x64 splitters—with loss levels of up to 21 dB.

In-Service Troubleshooting
The flexibility to add a singlemode live testing port for in-service troubleshooting enables you to fix problematic distribution fiber while the others are still active. Its filtered port blocks incoming transmission signals and its out-of-band wavelength will not interfere in your network. Maintaining an excellent quality of service (QoS), it will ease the troubleshooting testing steps.

Features FTB-730
• Test through high-port-count splitters (up to 1x128)
• Singlemode port for in-service troubleshooting with standard in-line OPM
• Dynamic range of up to 39 dB
• Short acquisition time to speed up deployment process
• EXFO Connect-compatible: automated asset management; data goes through the cloud and into a dynamic database
• iOLM-ready: one-touch multiple acquisitions, with clear go/no-go results presented in a straightforward visual format

Applications
• FTTx/MDU test challenges within PON networks
• Access network testing

Technical Specifications FTB-720:
• Model: FTB-730
• Wavelength (nm) b: 1310 ± 20/1490 ± 10/1550 ± 20/1625 ± 10/1650 ± 7
• Dynamic range at 20 ms (dB) c: 39/35/37/39 d / 37
• Event dead zone (m) e: 0.8
• Attenuation dead zone (m) e: 4 / 4.5 / 4.5 / 4.5 / 4.5
• Distance range (km): 1, 25, 2, 5, 5, 10, 20, 40, 80, 160, 260, 400
• Pulse Width (ns): 5, 10, 30, 50, 100, 275, 500, 1000, 2500, 10 000, 20 000
• Linearity (dB / dB) b: ± 0, 03
• PON dead zone (m) f: 35
• Loss threshold (dB): 0, 01
• Resolution Loss (dB): 0001
• Sampling resolution (m): 0, 04-5
• Sampling points: Up to 256 000
• uncertainties Distance (m) g: ± (0, 75 + 0, 0025% x distance + resolution)
• Time measurement: User-defined (. 60 minutes maximum)
• Typical real-time refresh (Hz): 4
• Stable source output power (dBm) h: -2.5
• Visual fault locator: n / a i
• Reflectance (dB) b: ± 2